News

 
  1. Receive Higher Accuracy when Measuring CTDIw

    06 Sep 10 RTI strongly recommends the 1-point method to be used for CTDIw measurements with the CT Dose Profiler. The method saves time and gives a higher accuracy than the 5 point method, either using a pencil ion chamber or the CT Dose Profiler.

  2. Decotron – New Distributor in Norway

    05 Sep 10 We welcome our new Norwegian distributor, Decotron, to our RTI Team.

  3. Meet Us at RPM in Bulgaria

    01 Sep 10 RTI will be at RPM in Bulgaria together with our distributor – Ecomed Services – to show you the best from RTI Electronics, our X-ray multimeters Piranha and Barracuda that togehter with our powerful software gives you an extraordinary X-ray QA solution.

  4. assets/News/Juli-2010/_resampled/SetWidth190-South-2.jpg

    RTI Electronics Strengthening its Position in South America

    23 Jul 10 We are now pleased to present three new distributors in South America.

  5. assets/News/Juli-2010/_resampled/SetWidth190-Picture-T20-034.jpg

    Important Solid State Detector Accessories by RTI Electronics

    23 Jul 10 If you own an RTI Piranha or Barracuda multi-meter, you may want to consider adding two special solid state detectors specially developed by RTI Electronics.

  6. EDGE July 2010

    21 Jul 10 Newsletter from RTI Electronics, July issue 2010.

Show all news

Dose Profile Analyzer

Dose Profile Analyzer

CT Dose Profile Analyzer Software


The software runs on a PC connected either to a Barracuda or a Piranha. After each scan (exposure), the data is presented instantly. The detector can be used for both point dose and CTDI measurements. Examples are available for measurements on different types of CT systems.
The picture below shows a CTDI measured in a head Phantom for a nominal beam width of 20 mm and pitch 0.969.

The dose rate profile has been acquired in the central position in the phantom. The cursors are automatically placed and centered at a distance of 100 mm for the calculation of the CTDI100.

Only one exposure is needed to measure and calculate CTDI100, CTDIw, CTDIvol, DLP, FWHM and Scatter index. To be able to simplify the procedure from normally 5 exposures to only one measurement in the central phantom position (CTDI100,c), a factor k is multiplied with CTDI100,c to get the CTDIw. The k factors are stored in the existing CT Database and have been acquired from a number of different CT systems.

In the software, you are also able to measure the CTDIw with the five exposures at different phantom positions but is seldom needed with the new "Central Point Method".

The software controls the Barracuda or Piranha. All settings for the meters and other pre-defined data are stored in the example files. Once a CT example has been loaded, measurement, storage and analysis of data can be done quickly for different types of CT scanners.
Measured data is saved and can be re-opened for later reviewing. Data can also be printed or exported to Microsoft Excel for further analysis.

 
 